11.1 LR8400-20, LR8401-20, LR8402-20 Memory HiLogger Specifications
230
(2) Measurement Functions
1: Settable for up to 15 channels (U1-1 to -15) when thermocouple burn-out is disabled.
2: Settable for up to 30 channels (up to U2-1 to U2-15) w
hen thermocouple burn-out disabled, or up
to 15 channels (U1-1 to U1-15) when enabled.
3: Settable for up to 60 channels (up to U3-1 to -15
and U4-1 to -15) when thermocouple burn-out
is disabled, or up to 30 channels (U2-1 to -15) when enabled.
Measurement
Recording interval 10ms
1
, 20ms
2
, 50ms
3
, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s,
30s, 1min, 2min, 5min, 10min, 20min, 30min, 1h
Recording length (time
span)
Set continuous recording Off or On.
When Off, any time can be set.
Repeating Measure-
ment Recording
Set Off or On.
When On, measurement repeats at the set recording interval.
Recording Data Quantity Analog recording n channels: 8 M/n data values
Waveform storage The most recent 8 million data values (one analog channel recording
time, or 8 M / n values for n channels) are stored in internal memory (n =
number of analog channels + pulse channels × 2 + digital channels × 1 +
alarm channels × 5 words)
Data retained in memory can be observed by backward scrolling.
Measurement data
backup
Data is retained for about 30 minutes after power-off
(except when powered off while writing
to removable storage, or within
five minutes after power-on)
Display
Display Screen Selec-
tion
(Hybrid display config-
urations)
• Simultaneous waveform, gauge and setting display
• Waveform display
• Simultaneous waveform and numerical value display
• Simultaneous numerical value and comment display
• Numerical value display
• Simultaneous waveforms and numerical calculation results display
• Waveform and A/B cursor values displayed at the same time
Waveform colors 24 colors
Waveform compres-
sion and magnification
Horizontal axis (time
axis)
100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min, 5min,
10min, 20min, 30min, 1h, 2h, 5h, 10h, 12h, 1day /divisions
Vertical axis (voltage
axis)
Select by position or upper/lower limits
• When selecting by position, magnification can be × 100, × 50, × 20, ×
10,
× 5, × 2, × 1, or × 1/2. 0 Position: Set between -50% and 150% (at
1X magnification)
• When selected by upper/lower limits, specify the limit values.
Waveform scrolling Time-axis scrolling is available by left/right arrow keys while measuring
and when measurement stops (waveform drawing period).
These keys make large jumps
Jump function Selects the displayed span of the waveform.