1.4 Basic Operation
24
See: About Continuous and Repeating Recording: (p. 50)
About Dead time
When repeating recording is enabled ([Repeat] set to [On])
After a recording length has elapsed, some “dead time” is needed for internal processing
before recording can resume. No measurement occurs during the dead time. So to avoid
information loss, split data into files at a specified interval by enabling continuous record-
ing ([Repeat]), and selecting [Split Save] for auto saving.
About Measuring Operation
Cont.
Repeat: Off Repeat: On
Off
Off
(Recording is
stop
ped before one
recording length
has elapsed)
On
(Repeat Recording: same as Off)
Stop
Measuring
Start
Measuring
Measuring
Recording time
Stop
Measuring
Start
Measuring
Measuring
Recording
time
Dead time
Stop
Measuring
Start
Measuring
Measuring
Recording time
Stop
Measuring
Start
Measuring
Measuring
Recording
time
Dead time
Stop
Measuring
Start
Measuring
Measuring
Measurement operation depends on whether the trigger or timing func-
tion is used (p. 117).