3.2 Configuring Measurement Settings
51
3
3
Chapter 3 Settings
Recording
Interval
(Interval)
Select the data acquisition interval.
When expansion input units are installed (as UNIT3 and UNIT4), the default
setting is 50 ms.
To select 10-ms intervals, first disable input units 2 to 4 (set to Off).
Repeat
Recording
(Repeat)
Select whether to automatically start a new measurement after stopping,
or to finish after one measurement.
Continuous
Recording
(Cont)
Select whether to measure for a specified time span, or continuously.
Recording
Time
(Record
time)
Set this when [Cont] is disabled (Off).
Specify the duration of one recording length (time span).
The maximum setting is limited by the recording capacity of internal
me
mory. The actual limit depends on the selected recording interval and
number of channels used for measurements, so the maximum settable
time is displayed as the [Max] value (indicated by the red box).
Setting options:( : default setting)
10ms, 20ms
, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min,
2min, 5min, 10min, 20min, 30min, 1h
Setting options:( : default setting)
Off
Measuring stop after one recording length (time span).
On When stopped by a stop trigger or elapsed recording time,
measurement automatically restarts.
Press ST
OP to stop measuring.
Repeated recording
On
Off
Setting options:( : default setting)
Off Normal recording (specify the recording time, and measure)
On
Select for continuous measurement.
Setting options:(default setting: 10 seconds)
d (0 to 499), h (0 to 23), min (0 to 59), s (0 to 59)
Change 1's digit
Change 10's digits
[Max], [Data pts]
To record for a longer time to removable storage de-
vice, enable continuous recording.
"Maximum Recordable Time" (p .12)