5.4 Performing Temperature Rise Test (Temperature Conversion Function (T))
117
5
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Enable the temperature conversion function. (T)
4
Set the initial resistance and initial temperature.
Set the initial resistance and initial temperature in Steps 1 through 3.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
Enables the function
Disables the function
(default) (go to step 6)
Selection
2
1
Setting range initial resistance : 0.001  to 9000.000 M (default: 1.0000 )
initial temperature: -10.0 to 99.9°C (default: 23.0°C)
Move the cursor to the setting you
wish to configure. Make the value
editable with the key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
The initial resistance value range varies with the scaling setting.