5.4 Performing Temperature Rise Test (Temperature Conversion Function (T))
118
Guideline for k
IEC 60034 recommends the following:
• Copper: k = 235
• Aluminum: k = 225
See: "Appendix 5 Temperature Conversion (T) Function" (p. A6)
5
Set the reciprocal (k) of the temperature coefficient at 0°C.
6
Return to the Measurement screen.
Setting range: -999.9 to 999.9 (default: 235.0)
Move the cursor to the setting you
wish to configure. Make the value