Chapter 13 Specifications
265
13
(18)Comparator Function
* When set using the instrument’s keys, the input range will reflect the range and scaling coefficient with a max-
imum resolution of 1 n and a maximum value of 9 G.
Operation Compares setting and measured values
Setting ON/OFF (fixed range when the comparator function is on; the comparator func-
tion is automatically turned off when the T and BIN functions are on)
Comparator mode ABS mode/ REF% mode
Default state OFF, ABS mode
Judgment Hi Measured value > Upper limit value
IN Upper limit value ≥ measured value ≥ Lower limit value
Lo Lower limit value > measured value
Total judgment function (RM3545-02 only)
Operation When using the MUX measurement terminal setting with the scan function set
to AUTO or STEP, a PASS/FAIL judgment is made for each channel, and a to-
tal judgment is determined.
PASS/FAIL judgment
(for each scan channel)
PASS When the comparator judgment satisfies the PASS conditions
FAIL When the comparator judgment does not satisfy the PASS condi-
tions
Total judgment PASS When all channels are PASS or when the PASS condition is OFF
FAIL When any channel is FAIL
PASS conditions OFF/ Hi/ IN/ Lo/ Hi or Lo/ ALL (for each scan channel)
Default setting IN
ABS Mode
Upper/Lower limit
ranges
0.000 0 m to 9000.00 M *
Default setting 0.000 0 m
REF% Mode
Display Absolute value display and Relative value display
Relative value display
range
-999.999% to 999.999%
Reference value range 0.000 1 m to 9000.00 M *
When using the MUX measurement terminal setting, the measurement results
for scan channel 1 can be used as the reference value. (RM3545-02 only)
Upper/ Lower limit
ranges
0.000% to ±99.999%
Default setting Reference value: 0.000 1 m, Upper/ Lower limit ranges: 0.000%
()
()
()
[]
%1001
基準値
測定値
相対値 ×
−=
(Relative value) =
(Measured value)
(Reference value)
× 100 [%]− 1