Chapter 13 Specifications
266
(19)BIN Function
* When set using the instrument’s keys, the input range will reflect the range and scaling coefficient with a max-
imum resolution of 1 n and a maximum value of 9 G.
(20)Comparator Beeper Setting
(21)Auto Hold Function
Operation Compares setting and measured values and displays the result.
Setting ON/ OFF (When the BIN function is on, the range and comparator functions are
fixed to off. When T is on while using the MUX measurement terminal setting,
the BIN function is automatically turned off.)
Comparator mode ABS mode/ REF% mode
Display Absolute value (resistance value) display only
BIN number 0 to 9
Default state OFF
Judgment Hi Measured value > Upper limit value
IN Upper limit value ≥ measured value ≥ Lower limit value
Lo Lower limit value > measured value
ABS Mode
Upper/Lower limit
ranges
0.000 0 m to 9000.00 M *
Default setting 0.000 0 m
REF% Mode
Reference value range 0.000 1 m to 9000.00 M *
Upper/ Lower limit
ranges
0.000% to ±99.999%
Default setting Reference value: 0.000 1 m, Upper/ Lower limit ranges: 0.000%
Operation Sounds a beeper based on the comparator judgment result or total judgment.
(Set separately for Hi/ IN/ Lo and for PASS/ FAIL when using the MUX mea-
surement terminals.)
Operation settings and
tones
type 1/ type 2/ type 3/ OFF
Number of beeps 1 to 5 times / continuous
Default setting OFF, 2times
Operation Holds measured values automatically (only when using the measurement ter-
minals on the front of the instrument with the internal trigger source and con-
tinuous measurement on [free-run]).
The hold is canceled when the measurement leads are removed from the tar-
get and the next measurement performed, or when the key is pressed.
Operation setting ON/ OFF
Default setting OFF