Chapter 13 Specifications
270
(3) Communications interfaces
Interface types GP-IB/ RS-232C/ PRINTER/ USB
Default setting RS-232C
RS-232C and printer communications settings
Communication
contents
Remote control, measured value output (export)
Transfer method Asynchronous, Full duplex
Transmission speed 9,600bps (default setting)/ 19,200bps/ 38,400bps/ 115,200bps
Data length 8 bit
Stop bit 1
Parity none
Delimiter Transmit: CR+LF, Receive: CR or CR+LF
Handshaking No X-flow, no hardware flow
Protocol Non-procedure
Connector Male 9-pin D-sub, with #4-40 attachment screws
USB
Communication
contents
Remote control, measured value output (export)
Connector Series B receptacle
Electrical specifications USB2.0 (Full Speed)
Class (mode) CDC Class (COM mode), HID Class (USB keyboard mode)
Default setting COM mode
Printer
Operation Prints data when the PRINT signal is input or when the print key is pressed.
Compatible printers Interface: RS-232C, no. of characters per line: 48 (single-byte) or more
Communication speed: 9,600bps/ 19,200bps/ 38,400bps/ 115,200bps
Data length: 8bit, Parity: none, Stop bit: 1bit, Flow control: none,
Message terminator (delimiter): CR+LF
Control codes: Must be able to print plain text directly.
Printing Contents Resistance measured values, temperature measured values, judgment re-
sults, measurement conditions, statistical results
Interval ON/ OFF
Interval time 0 to 3,600 s
Statistical calculations
clear
ON/ OFF
Number of columns
printed per row
1 column/ 3 columns
Default setting Interval: OFF, interval time: 1 s, statistical calculations clear: OFF,
number of columns printed per row: 1 column