Chapter 13 Specifications
271
13
GP-IB interface (RM3545-01 only)
Communication
contents
Remote control
Device address 0 to 31
Delimiter LF/ CR+LF
Default setting Device address: 1, Delimiter: LF
Miscellaneous Conforms to IEEE 488.2
Interface Functions SH1 All Source Handshake functions are supported.
AH1 All Acceptor Handshake functions are supported.
T6 Basic talker functions are supported.
Serial poll function are supported.
No talk-only mode.
The talker cancel function with MLA (My Listen Address) is supported.
L4 Basic listener functions are supported.
No listen-only mode.
The listener cancel function with MTA (My Talk Address) is supported.
SR1 All Service Request functions are supported.
RL1 All Remote/Local functions are supported.
PP0 No Parallel Poll function.
DC1 All Device Clear functions are supported.
DT1 All Device Trigger functions are supported.
C0 No Controller functions are supported.
Communications functionality
Remote function During remote operation by USB, RS-232C or GP-IB, all front panel key oper-
ations are disabled. Remote operation is canceled as follows:
LOCAL key, Reset, At power-on
By USB, RS-232C or GP-IB
:SYSTem:LOCal command
•By GP-IB
GTL command
Communications
monitor function
Displays the send/receive status of commands and queries.
Setting: ON/ OFF
Data output function During INT trigger source operation, measured values are output at TRIG sig-
nal or ENTER key input.
During EXT trigger source operation, measured values are automatically out-
put each time measurement completes.
(USB keyboard mode is available during INT trigger source use only.)
ON/ OFF
Memory function Measured values stored in the instrument’s memory are sent at once.
(The memory function is automatically turned off when using the MUX mea-
surement terminal setting.)
No. of memory: 50 (volatile memory, no backup)
ON/ OFF
Default setting Communications monitor function: OFF, Data output: OFF,
Memory function: OFF