Chapter 13 Specifications
272
(4) EXT I/O
Input Signals TRIG (IN0), CAL, KEY_LOCK, 0ADJ, PRINT (IN1), MUX, SCN_STOP,
LOAD0 to LOAD5
Valid only with BCD mode output: BCD_LOW
Optocoupler-isolated : no-voltage contact inputs (current sink/source output
compatible)
Input ON : Residual voltage; 1 V or less (Input ON current: 4 mA
(reference value))
Input OFF : OPEN (shutoff current: 100 A or less)
Response time : ON edge; Max. 0.1 ms, OFF edge; Max. 1.0 ms
Output Signals Output mode switching: JUDGE mode/ BCD mode
1. JUDGE mode: EOM, ERR, INDEX, HI, IN, LO, T_ERR, T_PASS, T_FAIL,
BIN0 to BIN9, OB, OUT0 to OUT2
2. BCD mode: EOM, ERR, IN, HILO
When BCD_LOW is ON : BCD1 to BCD3 × 4 digits,
RNG_OUT0 to RNG_OUT3
When BCD_LOW is OFF: BCD4 to BCD7 × 4 digits
Optocoupler-isolated : open-drain output (non-polar)
Maximum load voltage : 30 V DC
Residual voltage 1 V or less (load current: 50 mA)/
0.5 V or less (load current: 10 mA)
Maximum output current : 50 mA/ch
Default setting : JUDGE mode
Trigger Source Setting
Setting INT (Internal)/ EXT (External) (when using the MUX measurement terminals
with the scan function set to AUTO or STEP, fixed to EXT)
Default setting INT (Internal)
TRIG/PRINT filter function
Setting ON/ OFF
Operation During the response time, signal processing is performed only while the input
signal is held in the on state.
Response time 50 to 500 ms
Default setting OFF, 50 ms
TRIG Logic Setting
Setting OFF edge/ ON edge
Default setting ON edge