Selenia User Guide
Appendix BThe Mobile Selenia
132 MAN-03270 Revision 003
B.5 How to Test the System Integrity After Transport
Instructions: Photocopy this checklist for use after the system is relocated.
Name_____________________________________ Date_______________
System Serial Number____________________________________
Table 20: Integrity Checklist
Evaluated
Control Functions
Reference
Mechanical System Tests
Compression Up/Down
Controls and Functional Tests on page 33
Compression Release
Controls and Functional Tests on page 33
C-arm Rotation
Controls and Functional Tests on page 33
C-arm Up/down
Controls and Functional Tests on page 33
Collimator Override
Controls and Functional Tests on page 33
Light Field Lamp
Controls and Functional Tests on page 33
Smart Paddle System
Controls and Functional Tests on page 33
Emergency Off Switches
Emergency Off Switches on page 23
Quality Control Tests
Artifact Evaluation
Selenia QC Manual
SNR/CNR Measurement
Selenia QC Manual
Phantom Image
Evaluation
Selenia QC Manual
Compression Thickness
Selenia QC Manual