Selenia User Guide
Chapter 7—Maintenance and Cleaning
MAN-03270 Revision 003 111
7.2.2 How to Clean the Keyboard
Wipe the surfaces with a CRT wipe. If necessary, clean the keyboard with a vacuum. If
liquids enter the keyboard, contact Technical Support for a replacement.
7.3 Preventive Maintenance
Table 15: User Preventive Maintenance
Recommended Frequency
Maintenance Task
Description
Each Use Daily Weekly
Bi-
weekly
Monthly Quarterly
Semi-
annually
Clean & disinfect paddle
x
Clean & disinfect breast
platform
x
Visually inspect all paddles for
damage
x
Recommended/ required start
of operation verifications
x
x
All daily recommended/
required calibration
x
x
Diagnostic Review Workstation
Quality Control *
x
Detector Flat Field Calibration *
x
Artifact Evaluation *
x
Phantom Image *
x
Signal to Noise / Contrast to
Noise Measurements *
x
Compression Thickness
Indicator *
x
DICOM Printer Quality Control
x
View boxes and Viewing
Conditions
x
x
Repeat/Reject Analysis*
x
Compression *
x
Visual Checklist *
x
Emergency Stops
x
Clean array dust filter
x
* Refer to Quality Control Manual