Selenia User Guide
Chapter 7—Maintenance and Cleaning
MAN-03270 Revision 003 111
7.2.2 How to Clean the Keyboard
Wipe the surfaces with a CRT wipe. If necessary, clean the keyboard with a vacuum. If
liquids enter the keyboard, contact Technical Support for a replacement.
7.3 Preventive Maintenance
Table 15: User Preventive Maintenance
Recommended Frequency
Maintenance Task
Description
Each Use Daily Weekly
Bi-
weekly
Monthly Quarterly
Semi-
annually
Clean & disinfect breast
x
Visually inspect all paddles for
damage
Recommended/ required start
of operation verifications
All daily recommended/
required calibration
Diagnostic Review Workstation
x
Detector Flat Field Calibration *
Signal to Noise / Contrast to
Noise Measurements *
Compression Thickness
Indicator *
DICOM Printer Quality Control
View boxes and Viewing
Conditions
* Refer to Quality Control Manual