Selenia User Guide
Chapter 4—Startup, Functional Tests, and Shutdown
38 MAN-03270 Revision 003
4.3 Monthly System Tests
• Test all Emergency Stop Switches. Follow the on-screen reset procedure.
• Change the AEC Sensor Position switch, and make sure the compression device
indicator and the Acquisition Workstation display change.
• Test all buttons on all keypads and footswitches. Include the compression release on
the Acquisition Workstation.
• During Compression Device tests make sure the thickness and force displays change.
• During C-arm rotation tests, confirm that the LED Angle displays change.
4.4 How to Restart the System
4.4.1 Restart From Sleep Mode
1. To activate the system, move the trackball or mouse.
2. In the Login dialog box, enter reboot for the user name and password.
3. Wait for the system to restart.
4. Log on to the Operating System, then Log on to the Application software.
5. If the system was off for less than 30 minutes, select Dismiss in the Launch dialog
box.
4.4.2 Restart After the Emergency Off Switch was Activated
1. Turn the Emergency Off switch by one-quarter turn clockwise to reset the switch.
2. When the Power Panel dialog box displays, "Communication with the generator is
lost," examine the Emergency Off Switches and the circuit breaker:
Caution:
Do not close this dialog box by using the X in the upper
corner.
• If the Gantry power is turned on, select the Cancel button.
• If the Gantry power is turned off, and the Emergency Off Switches and the circuit
breaker are set correctly, select the OK button.
3. When the "Link established with generator" dialog box appears, select OK.