Selenia User Guide
Chapter 2—General Information
MAN-03270 Revision 003 9
Chapter 2
Chapt er 2 General Information
2.1 Warnings and Precautions
Read and understand this manual before you use the system. Always follow all the
instructions in this manual.
This system is for use by qualified Operators only. The Operators must have experience in
the principles of mammography before following the instructions contained in this
manual.
This system is classified as CLASS I, TYPE B APPLIED PART, IPX0, permanently
connected equipment, continuous operation with short term loading per IEC 60601-1.
There are no special provisions to protect the system from flammable anesthetics or
WARNING!
Risk of electric shock. Only connect this
equipment to supply mains with Protective
Earth.
WARNING!
Only trained Service Engineers authorized
through Hologic can open any of the panels.
This system contains lethal voltages.
WARNING!
Electrical equipment used near flammable
anesthetics can cause an explosion.
WARNING!
After power failure, remove the patient from the
system before you apply power.
WARNING!
To correctly isolate the system, attach only
approved accessories or options to the system.
Only approved personnel can change the
connections.