Selenia User Guide
Chapter 2General Information
10 MAN-03270 Revision 003
WARNING!
Keep a 1.5 meter safe distance between the
patient and any non-patient devices.
Do not install non-patient system components
(like the Workflow Manager, the diagnostic
review workstation, or the hard copy printer) in
the Patient Area.
WARNING!
For North American electrical safety
requirements, use a Hospital Grade receptacle to
supply a correct Ground.
Warning:
The user or a servicing engineer must correct
problems before the system is used.
Warning:
The user must prepare for preventive maintenance
by an approved servicing engineer.
Warning:
This system can be dangerous to the patient and the
user. Always follow the safety precautions for x-ray
exposures.
Warning:
The disk drives installed in this system are a Class I
Laser Product. Prevent direct exposure to the beam.
Hidden laser radiation exists if the case to a disk
drive is open.
Warning:
Only qualified users can use this system.
Warning:
Control the access to the equipment according to
local regulations for radiation protection.
Warning:
Keep your full body behind the radiation shield
during the exposure.