Selenia User Guide
Chapter 2General Information
MAN-03270 Revision 003 11
Warning:
This device contains dangerous material. Send
decommissioned material to Hologic or contact your
service representative.
Warning:
The equipment has motors. You must be careful
when you adjust the equipment for patient use.
Observe equipment and patient at all times during
setup. If a chair is necessary, use an adjustable chair
set above its minimum height.
Warning:
Put the footswitches away from the patient and
C-arm area to prevent any accidental footswitch
operation. When the patient has a wheelchair, put the
footswitches away from the area.
Warning:
Do not leave the patient during the procedure.
Warning:
Keep the hands of the patient away from all buttons
and switches at all times.
Warning:
You increase the patient dose to high levels when
you increase the AEC exposure adjustment. You
increase the image noise or decrease image quality
when you decrease the AEC exposure adjustment.
Warning:
If a paddle touches possible infectious materials,
contact your Infection Control Representative to
remove contamination from the paddle.
Warning:
The bar code scanner installed in this system is a
Class II Laser Product. Prevent direct exposure to the
beam. Hidden laser radiation exists if the cover is
opened.