Selenia User Guide
Chapter 2—General Information
12 MAN-03270 Revision 003
Warning:
The Face Shield does not protect the patient from
radiation.
Caution:
Do not use any hot source (like a heating pad) on the
image receptor.
Caution:
Never turn off the Acquisition Workstation or Value
Console Circuit Breaker except in emergency. The circuit
breaker can turn off the Uninterruptible Power Supply
(UPS) and risk data loss.
Caution:
To prevent possible damage from thermal shock to the
Digital Image Receptor, follow the recommended
procedure to turn off the equipment.
Caution:
Risk of data loss. Do not put any magnetic media near or
on devices that create any magnetic fields.
Caution:
The system is a medical device and not a normal computer.
Only make approved changes to the hardware or software.
Install this device behind a firewall for network security.
The computer virus protection or network security for this
medical device is not supplied (for example, a computer
firewall). The network security and anti-virus provisions
are the responsibility of the user.
Caution:
To prevent errors and possible data loss, only use
approved accessories with this equipment.