Selenia User Guide
Chapter 6—The User Interface
66 MAN-03270 Revision 003
6.2.7.2.6 AEC Alarm Messages
When the system cannot use the selected AEC mode to acquire an Image, one of the
following messages displays.
Reason and Corrective Action
Time Less Than Min
The AEC Sensor is over a thin or small area of breast tissue.
Use the AEC Sensor in a position under the breast tissue.
Alternately, use AutoTime and a lower kV.
Time Exceeds Max
Very dense breast tissue, an implant, pacemaker, or other
anatomy changed the AEC sensor. Move the manual AEC
sensor where there are no obstructions. Alternately, use
AutoTime and a higher kV.
1. Select OK in the dialog box.
2. Select the Reset Generator button.