EasyManua.ls Logo

Honeywell SmartLine ST 700 - Optional Integrated Two-Button Assembly (Standard Display)

Honeywell SmartLine ST 700
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision 13 ST 700 SmartLine Pressure Transmitters User’s Manual Page 7
2.6 Optional Integrated Two-Button Assembly (Standard Display)
The standard display does not support all the transmitter configuration parameters and has limited
features.
The optional Integrated two-button assembly for the standard display provides the following features
and capabilities:
Menu and enter key functionality.
With the menu-driven display:
o Comprehensive on-screen menu for navigation.
o Transmitter configuration: enter LRV, enter URV and loop test.
o Transmitter calibration.
o Display configuration (contrast only).
o Set zero and span parameters.
If you are using the optional external two-button assembly with the
standard display then you can perform all the above operations without
removing external glass cap using the external buttons.
Lower range value (LRV): A display parameter (standard display), which
allows users to enter the measuring value for which the analog output will
be scaled to 4mA.
Upper range value (URV): A display parameter (standard display), which
allows users to enter the measuring value for which the analog output will
be scaled to 20mA.

Table of Contents