62 ST 800 Pressure Transmitter User’s Manual Revision 15
Линейн (LINEAR)
КвдрКор (SQRT)
Pressure Transfer function:
Select the pressure transfer
function for calculated output:
Linear: (output scaled linearly
per configured range) Square
Root: (output scaled as square
root conversion per
configured range – for flow
measurements)
Press to
enter menu
selection
Menu to
select from
list
to enter
Нолтечь (ZROFLOW)
1Отсечк (SBRKPT)
2Отсечк (DUALSLP)
Low Flow Cutoff Mode:
For Square Root (Transfer
function) configurations only,
choose the operating mode to
determine the low flow cutoff
breakpoint: Dual slope:
(breakpoint determined by dual
slope calculation) User
configuration: (breakpoint
defined by user)
Zero Flow: Sets analog output
to 4mA when square root /
flow is below user defined
flow cutoff value.
Note: Sets loop output to low
saturation when the differential
pressure is below
–10%Flow.
Press to
enter menu
selection
Menu to
select from
list
to enter
When “User Configuration” is
selected for Low Flow Cutoff
Mode, enter the desired
breakpoint value.
Press to
enter menu
selection
to select
number.
to enter and
shift to the
next digit to
the right
SP Zero Correct
[24SPZRO]
Corrects the static pressure to a
zero value psi for DP transmitters
only
Performs the correction of
measured static pressure value
to a value of zero psi.
Executing this selection
corrects Static Pressure Zero
based on the input static
pressure.
The live value of the static
pressure input is shown on this
display so the user can
easily see the result of the
correction.
Press to
enter menu
selection
Press to
initiate action