34 tirb45e1 (2019-02) CDGxxxDxx.cp
A device reset is initiated when the following byte sequence is sent.
Command
Byte 0
Byte 1
Byte 2
Byte 3
Byte 4
0x74
0x65
0x73
0x65
0x72
Status (supported values)
0
1
2
3
255
Last command completed, no error, no response
Reserved
Last command completed, error, no response
Reserved
Command is executing
Response
Byte 0
See Subindex 0x02
Byte 1
not used = 0x00
Execution of this command will store all parameters to non-volatile memory. If a
device automatically saves all non-volatile parameters at the time they are set, this
command will not take any action.
Index
SI
DataType
NV
Access
PM
Name
FBF2
0x01
V_STRING(4)
RW
Command
0x02
BYTE
RO
Status
0x03
V_STRING(2)
RO
Response
A device reset is initiated when the following byte sequence is sent.
Read:
Command: Read
Byte 0
Byte 1
Byte 2
Byte 3
0x01 = slave saves the non-volatile parameters
when writing 0xFBF2:01 with 0x65766173
not used = 0x00
not used = 0x00
not used = 0x00
Write:
Command: Write
Byte 0
Byte 1
Byte 2
Byte 3
0x73
0x61
0x76
0x65
Status (supported values)
0
1
2
3
255
Last command completed, no error, no response
Reserved
Last command completed, error, no response
Reserved
Command is executing
Subindex 0x01
Subindex 0x02
Subindex 0x03
4.7.26 Store Parameters
Command
Subindex 0x01
Subindex 0x02