40    tirb45e1 (2019-02) CDGxxxDxx.cp 
 
  0x03  BOOL    RW    Alarm Enable 
 
 
Specifies the behavior for the value for States other than Valid. 
 
0 
1 
2 
Zero 
Full Scale 
Hold last value 
 
 
Safe Value: The value to be used for Safe State = Safe Value. 
 
 
Alarm Enable: Enables the setting of the Alarm Status Bits. 
 
 
Warning Enable: Enables the setting of the Warning Status Bits. 
 
 
Alarm Trip Point High: Determines the value above which an Alarm Condition will 
occur. 
 
 
Alarm Trip Point Low: Determines the value below which an Alarm Condition will 
occur. 
 
 
Alarm Hysteresis: Determines the amount by which the value must recover to clear 
an Alarm Condition. 
 
 
Warning Trip Point High: Determines the value above which a Waning Condition 
will occur. 
 
 
Warning Trip Point Low: Determines the value below which a Waning Condition will 
occur. 
 
 
Warning Hysteresis: Determines the amount by which the value must recover to 
clear a Warning Condition. 
 
 
 
 
 
 
5.3.2  Manufacturer 
5.3.3  Manufacturer