1 - 16
Transpector XPR 3+ Operating Manual
Figure 1-13 Cable and connector elbow feed
6 Connect the red and black wires to terminals 1 and 2 in either order. This can
be done by soldering the wires in place or using the appropriate screw
terminals on the connector (3) assembly.
Figure 1-14 Connecting the red and black wires to the terminal
7 Press the connector (3) back into the connector elbow (4) screw the nut (5)
securely into place. Connect the connector (3) to the solenoid coil (1) noting the
connection scheme NOTE: When fitting the connector (3) into the connector
elbow (4) note the orientation of the connection to the solenoid coil (1). Ideally
the elbow should be facing away from the inlet air connection point as it can
interfere with the connection of the inlet air.