Checking the timer for the second phase
7. Tap the TIMER key to get to the time settings for the second
phase (
Time 2
).
or 00:00 if no time has been entered.
In the separate timer display area the signs
Timer
and
Phase 2
are lit up.
8. Make sure the second timer is turned off. If not, use the
ON/OFF key in the lower setting key area to turn the timer
(
Time 2
) off.
Entering the setpoints for the second phase and switching on
the parameters
After making sure, that the second timer (
Time 2
) is switched off, the
different parameters for phase 2 can be selected using the Select key
and afterwards be switched on.
INFORMATION
If the timer should be used to stop the cultivation process after a
defined time, all parameters for phase 2 can be switched off. This
causes the parameters to switch off after the set time for phase 1
has elapsed and therefore to stop the cultivation process.
9. Tap the Select key to select the desired parameters.
In both display areas the first integral pair of parameters (
Temp
and
RPM
) appears with the last setpoint values entered or the
message
OFF
.
The sign
Setpoint
indicates, that the setpoints can be entered.