CUSA® Clarity System Service and Maintenance Manual Page 50 of 117 0645403-2-EN
Handling and Transportation of the System
If you need to move/relocate the CUSA Clarity Ultrasonic Surgical Aspirator System within the environment of the Healthcare Facility,
note the guidelines below:
Before moving the console:
• Power off the system.
• Unplug the power cord from the wall receptacle and secure it.
• Disconnect the handpiece.
• Disconnect and remove the irrigation fluid bag.
• Lower the IV pole.
• Secure any loose objects. Make sure there are no loose objects placed on the top of the console.
• Secure the footswitch and cord.
When moving the console (if aached to a cart):
• Push the cart using the handle.
• Do not run while pushing the console and cart.
• Use an elevator to move the console between floors of a building. Never use a stairwell.
• Do not aempt to li the console while aached to the cart.
Storage of the System and Accessories
Console
• Make sure the system is clean. Store the system in a low traffic area that is free of dirt, blood, water, and other contaminants.
• Contamination guard may be stored with console if it is not expired.
• Store the system at an ambient temperature between -20°C (-4°F) and 60°C (154°F).
Console with Cart
• Wrap footswitch cord around cart and store footswitch in cart basket.
• Wrap power cord around cart.
• Make sure the system is clean. Store the system in a low traffic area that is free of dirt, blood, water, and other contaminants.
• Contamination guard may be stored with console if it is not expired.
• Store the system at an ambient temperature between -20°C (-4°F) and 60°C (154°F).
Handpiece
• Store the handpiece and torque base in the sterilizer case according to your facility’s policy.
• Store the system at an ambient temperature between -20°C (-4°F) and 60°C (154°F).