EasyManua.ls Logo

Integra CUSA Clarity System - Handpiece; Electrical Requirements; Power Input and Fusing; Power Cords

Integra CUSA Clarity System
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
This document contains proprietary and confidential information of Integra LifeSciences Corporation. Integra’s confidential information may not be used, disclosed or reproduced without the prior wrien consent of Integra LifeSciences Corporation.
CUSA® Clarity System Service and Maintenance Manual Page 89 of 117 0645403-2-EN
10.5 Handpiece
Dimensions
Length 13.03 cm (5.13 in)
Diameter 2.41 cm (0.95 in)
Weight 71 g (0.15 lb)
Cable Length 463.3 cm (182.4 in)
Frequency
High 36 kHz
10.6 Electrical Requirements
10.6.1 Power Input and Fusing
Voltage 100-240V~
Frequency 50-60Hz
Power 480VA
Phases 1
Class 1
Fuses 5A, Type T, 1500A IR, 250V, 5mm x 20mm, ceramic body
10.6.2 Power Cords
United States and Canada:
Meets hospital grade UL817
Cord 3 x 16AWG 5 Meters Color White
Console Connector C13 IEC 60320-1 V-Lock
Plug NIMA 5-15 Straight
Other Countries:
Cord
3 x 1mm2 5 Meters Color Black
Console Connector C13 IEC 60320-1 V-Lock
NOTE
Only the exact fuse type and rating is to be used. Lielfuse 0215005.HXP, 2 fuses required per console.

Table of Contents