CUSA® Clarity System Service and Maintenance Manual Page 52 of 117 0645403-2-EN
Frequency Tas k Procedure
Upon receipt of System Cleaning Section 6.3 Cleaning and Sterilizing
Upon receipt, following cleaning Visual Inspection Section 7.7 Visual Inspection
Upon receipt, following cleaning Perform functional tests upon receipt
of the console from the distributor –
troubleshoot as required.
Chapter 7: Functional Testing
Section 8.2: Exporting Logfiles
Every six months, when the color changes, or
when the surgeon notices a lack of suction.
Replace the contamination guard. Section 5.6 Connecting the Tubing Set
Annually Preventive Maintenance Contact service for assistance
Annually Replace PC lithium baery. Contact service for assistance
Annually Replace Suction motor pump controller
(S30904193).
Contact service for assistance
Annually Replace Contact, Grounding Finger
(72905223)
Contact service for assistance
Every 4 years Replace 8GB mSATA card (72904462). Contact service for assistance
Table 7-2 Periodic Maintenance Schedule
Ordering Replacement Parts
Order the following replacement parts from an Integra Service Center:
• IV Pole
• Power Cord
• Footswitch
• Contamination Guard
When ordering replacement parts for equipment, include the following information:
• Model number (located on the CUSA Clarity console rear panel)
• Serial number (located on the CUSA Clarity console rear panel)
7.6 Periodic Maintenance Schedule
The CUSA Clarity system is designed to meet the requirements specified in IEC 62353, Edition 2.0 2014-9. Record all periodic maintenance activities on
the data sheet supplied with this manual.