Plug variation by country:
Australia AS3112, Straight
Brazil NBR 14136, Straight
China GB2099, Straight
Europe CEE 7/ XVII, Straight
Italy CEI 23-50, Straight
Japan JIS 8303, Straight
South Africa SANS 164-1, Straight
Switzerland SEV Typ 12, Straight
United Kingdom BS 1363, 13A fused, Angled
Only use Integra approved power cords.
10.7 Duty Cycle
10.6.3 Maximum Low Frequency Leakage
Touch current <500µA
Patient leakage current <50µA
Under maximum loading conditions, the CUSA Clarity console is suitable for ultrasonics activation times of 10 minutes on, 5 minutes off.
The system requires a minimum of one hour exposure at its operating temperature range before you use it.
See the EMC Information section, section 10.10 for further information on environmental conditions.
Electromagnetic Interference
The CUSA Clarity system console minimizes electromagnetic interference to other equipment used in the operating room. The system complies
with the requirements of IEC 60601-1-2: 2014 Fourth Edition.
WARNING
Explosion Hazard - Do not use the CUSA Clarity System in the presence of flammable anesthetics or other volatile solvents.
WARNING
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.