CUSA® Clarity System Service and Maintenance Manual Page 70 of 117 0645403-2-EN
8.10 Error Code Matrix
For all error codes, except those explained below, contact Integra Service and Repair for assistance..
Error Code Description Root Cause Solution 1 Solution 2
E1208 Vacuum Pressure Low Vacuum is lower than expected Connect a contamination
guard
Fix leakage on suction
tubing set
E1220 Irrigation Motor RPM Irrigation motor speed is different
from the demand. Check if something
is blocking the irrigation pump head
Fix blockage at the pump
head
Call Service for assistance
E1226 Pinch Solenoid Position Pinch valve is not at the expected
position (staying opened)
Check for blockage Call Service for assistance
E1400 Pinch Solenoid Position Pinch valve is not at the expected
position (staying closed)
Check for blockage Call Service for assistance
E2201 Broken footswitch microswitch
failure
Microswitch from Ultrasonic
footswitch is defective
Replace footswitch Call Service for assistance
E2202 Broken footswitch potentiometer Potentiometer from Ultrasonic
footswitch is defective
Replace footswitch Call Service for assistance
E2203 Broken fast flush footswitch Microswitch from fast flush footswitch
is defective
Replace footswitch Call Service for assistance
E2303 Invalid handpiece EEPROM version Handpiece EEPROM corrupted Replace handpiece Call Service for assistance
E2304 Invalid CRC on handpiece EEPROM Handpiece EEPROM corrupted Replace handpiece Call Service for assistance
E2305 Handpiece EEPROM parameter out
of range
Handpiece EEPROM corrupted Replace handpiece Call Service for assistance
Table 8-6