4 Imaging programs
Exposure settings for 3D imaging, medium panel (default values)
NOTICE! Voltage setting is always 90kV with the unit 3D modality.
FOV Resolution Exposure
time
Scan
time
Voxel
size
(um)
Default
mA
mAs DAP
(mGycm2)
Low Dose 1,2 s 11,0 s 280 3,2 3,8 32
Std Res 2,3 s 11,0 s 200 8 18,4 162
High Res 6,1 s 17,4 s 125 6,3 38,4 332
50 x 50
mm
Endo Res 8,7 s 17,4 s 85 6,3 54,8 474
Low Dose 1,2 s 11,0 s 320 3,2 3,8 58
Std Res 2,3 s 11,0 s 300 8 18,4 288
61 x 78
mm
High Res 6,1 s 17,4 s 200 6,3 38,4 591
Low Dose 1,2 s 11,0 s 320 3,2 3,8 72
Std Res 2,3 s 11,0 s 300 8 18,4 358
78 x 78
mm
High Res 6,1 s 17,4 s 200 6,3 38,4 735
Low Dose 2,3 s 21,1 s 400 3,2 7,4 138
Std Res 4,5 s 29,5 s 350 8 36,0 689
78 x 150
mm
High Res 8,1 s 33,3 s 250 6,3 51,0 983
Low Dose 4,5 s 42,2 s 420 3,2 14,4 276
Std Res 8,1 s 38,2 s 380 5 40,5 780
130 x
150 mm
(optional)
High Res 8,1 s 38,2 s 320 8 64,8 1248
Default exposure settings for scout imaging (3D)
FOV Resolution kV mA Exposure time
61 x 41 mm Scout 90 12.5 0,02 s
61 x 78 mm Scout 90 12.5 0,04 s
Default exposure settings for scout imaging
FOV Resolution kV mA Exposure time
50 x 50 mm Scout 90 12.5 0,02 s
61 x 78 mm Scout 90 12.5 0,02 s
78 x 78 mm Scout 90 12.5 0,02 s
78 x 150 mm Scout 90 12.5 0,04 s
130 x 150 mm
(optional)
Scout 90 12.5 0,04 s