9 Technical specifications
KaVo Scan eXam One 69
INSTALLATION REQUIREMENTS &
ENVIRONMENT CONTROL:
In order to minimize interference risks, the
following requirements shall apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must
meet the requirements given in Technical
specifications. Use of incorrect cables may result in
the device causing radio frequency interference.
Electrostatic discharges environment &
recommendations
In order to reduce electrostatic discharge
interference, a charge dissipative floor should be
installed to prevent charge accumulation.
• The dissipative floor material must be
connected to the system reference ground, if
applicable.
• Relative humidity must be maintained above 30
percent.
Stacked components & equipment
WARNING! The device should not be used
adjacent to or stacked with other equipment; if
adjacent or stacked use is necessary, the device
should be observed to verify normal operation in
the configuration in which it will be used.
Interference may occur in the vicinity of equipment
marked with the following symbol: