17-8 SCPI Reference Tables
Command Description
Default
parameter
Ref
SCPI
:LGRoup <name> Specify autorange limit; HIGH or LOW. HIGH
:LGRoup? Query upper limit for autorange.
:AUTO? Query state of autorange.
:ADIScharge Path for auto discharge: Sec 5
:LEVel <NRf> Set auto discharge level; -21e-6 to 21e-6. 2e-6
:LEVel? Query auto discharge level.
[:STATe] <b> Enable or disable auto discharge. OFF
[:STATe]? Query state of auto discharge.
:AVERage Path to control the Digital Filter: Sec 6
:TCONtrol <name> Select filter control; MOVing or REPeat. REP
:TCONtrol? Query filter control.
:COUNt <n> Specify filter count; 2 to 100. 10
:COUNt? Query filter count.
[:STATe] <b> Enable or disable digital filter. OFF
[:STATe]? Query state of digital filter.
:MEDian Path to control median filter: Sec 6
:RANK <NRf> Specify “n” for rank; 1 to 5 (rank = 2n+1). 1
:RANK? Query rank.
[:STATe] <b> Enable or disable median filter. OFF
[:STATe]? Query state of median filter.
Note: *RST default is ON and SYSTem:PRESet default is OFF.
Table 17-4 (cont.)
SENSe command summary

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals