SCPI Reference Tables 17-11
Table 17-7
SYSTem command summary
Command Description
Default
parameter
Ref SCPI
:SYSTem Sec 16
:ZCHeck <b> Enable or disable zero check. ON Sec 2
:ZCHeck? Query state of zero check.
:ZCORrect Zero correct: Sec 2
[:STATe] <b> Enable or disable zero correct. OFF
[:STATe]? Query state of zero correct.
:ACQuire Acquire a new zero correct value.
:PRESet Return to SYSTem:PRESet defaults. √
:LFRequency <freq> Select power line frequency; 50 or 60 (Hz). Sec 1
:LFRequency?
:AZERo Path to control autozero: Sec 2 √
[:STATe] <b> Enable or disable autozero. ON √
[:STATe]? Query state of autozero. √
:TIME Timestamp:
:RESet Reset timestamp to 0 seconds.
:POSetup <name> Select power-on setup; RST, PRESet, or SAVx
(where x = 0 to 4).
:POSetup? Query power-on setup.
:VERSion? Query SCPI revision level. √
:ERRor Read error queue: (see Note ) Sec 13 √
[:NEXT]? Read and clear oldest error/status (code and
message).
:ALL? Read and clear all errors/status (code and
message).
:COUNt? Read the number of messages in queue.
:CODE Code numbers only:
[:NEXT]? Read and clear oldest error/status (code only).
:ALL? Read and clear all errors/status (codes only).
:CLEar Clear messages from error queue. Sec 13
:KEY <NRf> Simulate key-press; see Figure 16-3. √
:KEY? Query the last “pressed” key. √