17-14 SCPI Reference Tables
Command Description
Default
parameter
Ref SCPI
:TRIGger:CLEar Clear input triggers immediately.
[:TCONfigure]
:DIRection <name> Enable (SOURce) or disable (ACCeptor)
bypass.
ACCeptor
:DIRection? Query trigger source bypass.
[:ASYNchronous] Configure input/output triggers:
:ILINe <NRf> Select input trigger line; 1, 2, 3, 4, 5, or 6. 1
:ILINe? Query input trigger line.
:OLINe <NRf> Select output trigger line; 1, 2, 3, 4, 5, or 6. 2
:OLINe? Query output trigger line.
:OUTPut <name> Output trigger after measurement (SENSe) or
not at all (NONE).
NONE
:OUTPut? Query output trigger status.
Table 17-9 (cont.)