EasyManua.ls Logo

Keithley SourceMeter 2410

Keithley SourceMeter 2410
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2400 Series SourceMeter
®
User’s Manual Sweep Operation 10-31
Call OutputCmd(intGPIB, “*RST”) 'Restore GPIB default
conditions.
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”)
Call OutputCmd(intGPIB, “:SENSE:FUNC:CONC OFF”) 'Turn off Concurrent
Functions.
Call OutputCmd(intGPIB, “TRIG:COUN 3”) 'Trigger count = #
sweep points.
Call OutputCmd(intGPIB, “:SOUR:FUNC MEM”) 'Source Memory sweep
Mode.
Call OutputCmd(intGPIB, “:SOUR:MEM:POIN 3”) 'Number of Memory
Points = 3.
Call OutputCmd(intGPIB, “:SOUR:MEM:STAR 1”) 'Start @ Memory Loca-
tion 1.
Call OutputCmd(intGPIB, “SOUR2:CLE:AUTO ON;:SOUR2:CLE:AUTO:DEL 1”)
'Set Digital I/O
auto-clear
Call OutputCmd(intGPIB, “:CALC2:cLIM:PASS:SOUR2 12” 'Set Digital Output
Pattern for Limit #2
Passing
Call OutputCmd(intGPIB, “OUTPUT ON”) 'Turn on Output
Call OutputCmd(intGPIB, “:INIT”) 'Trigger Sweep.
End Sub