EasyManuals Logo

Keithley SourceMeter 2410 User Manual

Keithley SourceMeter 2410
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #561 background imageLoading...
Page #561 background image
F-8 Contact Check Function 2400 Series SourceMeter
®
User’s Manual
Figure F-5
Grading mode contact check limit testing
Perform
Limit 1 Test
?
Yes
No
Pass
?
No
Yes
Display
“FAIL”
Perform
Limit 2,3,5-12
Tests
?
Yes
No
Pass
?
No
Yes
Another
Test Cycle
?
No
Yes
Start
Turn Output ON and press LIMIT key.
Output First
Fail Pattern
Test
Another
Device
?
No
Yes
Stop
Press LIMIT key
Yes
No
Wait for SOT pulse
from handler
Perform Source-
Measure action
Display
“FAIL”
Binning
Control
Binning
Control
First
Failure
?
First
Failure
?
End
End
Yes
Yes
Output Limit 1
Fail Pattern
Immediate
Output Limit 2,3,5-12
Fail Pattern
Immediate
Store Limit 1 Fail
Pattern in Memory
Store Limit 2,3, 5-12 Fail
Pattern in Memory
No
No
Any
Failures
?
Display
“PASS”
Yes
Binning
Control
End
Any
Failures
?
Immediate
Output Pass
Pattern
No
Perform
Limit 4 Test
?
Yes
No
Pass
?
No
Yes
Store Limit 4 Fail
Pattern in Memory
No
First
Failure
?
Yes
Binning
Control
End
Immediate
Output Limit 4
Fail Pattern
Display
“FAIL”
Perform
Contact Check
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals