2400 Series SourceMeter
®
User’s Manual Status Structure 15-11
Status register sets
As shown in Figure 15-1, there are four status register sets in the status structure of the
SourceMeter; Standard Event Status, Operation Event Status, Measurement Event Status,
and Questionable Event Status.
NOTE See Appendix B for details on which register bits are set by specific error and
status conditions.
Register bit descriptions
Standard Event Register
The used bits of the Standard Event Register (shown in Figure 15-4) are described as follows:
Bit B0, Operation Complete — Set bit indicates that all pending selected device
operations are completed and the SourceMeter is ready to accept new commands.
This bit only sets in response to the *OPC? query command. See Section 16 for
details on *OPC and *OPC?.
Bit B1 — Not used.
Bit B2, Query Error (QYE) — Set bit indicates that you attempted to read data
from an empty Output Queue.
Bit B3, Device-Dependent Error (DDE) — Set bit indicates that an instrument
operation did not execute properly due to some internal condition.
Bit B4, Execution Error (EXE) — Set bit indicates that the SourceMeter detected
an error while trying to execute a command.
Bit B5, Command Error (CME) — Set bit indicates that a command error has
occurred.
Command errors include:
IEEE-488.2 syntax error — SourceMeter received a message that does not fol-
low the defined syntax of the IEEE-488.2 standard.
Semantic error — SourceMeter received a command that was misspelled or
received an optional IEEE-488.2 command that is not implemented.
The instrument received a Group Execute Trigger (GET) inside a program
message.
Bit B6, User Request (URQ) — Set bit indicates that the LOCAL key on the
SourceMeter front panel was pressed.
Bit B7, Power ON (PON) — Set bit indicates that the SourceMeter has been
turned off and turned back on since the last time this register has been read.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals