Page #233 background image
10-14 Sweep Operation 2400 Series SourceMeter
®
User’s Manual
SML 007 — Leakage Current Test
Source +V, Measure I.
Limit 2 test – Min/max limits for current reading.
Summary – This test is the same as the test at memory location 004, except the
source voltage is reversed to properly bias the diode that was installed backwards.
Test results — The test results for the 4-point source memory sweep are stored in the
buffer. Stored readings are accessed by pressing the RECALL key. Each of the four read-
ing numbers are preceded by a “P” or an “F” to indicate the “Pass” or “Fail” result of the
corresponding test. See Section 9 for details on the data store.
Configuring and running a sweep
Front panel sweep operation
Sweep configuration menu
The sweep configuration menu is structured as follows and is shown in Figure 10-7. Note
that bullets indicate the primary items of the sweep menu and dashes indicate the options
of each menu item. Using Section 1, “Rules to navigate menus,” go through the following
menu to select and configure the desired sweep.
CONFIGURE SWEEPS menu:
Press CONFIG then SWEEP to display the sweep configuration menu.
TYPE — Use this menu item to select the type of sweep:
–STAIR — When the linear staircase sweep is selected, you will be prompted to
enter the START, STOP, and STEP levels.
LOG — When the logarithmic staircase sweep is selected, you will be
prompted to enter the START and STOP levels and specify the number of
measurement points.
CUSTOM — With the custom sweep selected, you specify the number of
measurement points (# POINTS) in the sweep and the source level at each
point (ADJUST POINTS). With the INIT option, you can set a consecutive
range of measurement points in the sweep to a specific level. For example,
assume that for a 20-point custom voltage sweep (# POINTS = 20), you want
points 10 through 15 to be set for 1V. After selecting the INIT option, set the
VALUE to +1.000000V, set the START PT to 10, and set the STOP PT to 15.
SRC MEMORY — With the Source Memory Sweep selected, you specify the
memory location START point to start the sweep (1 is the default) and the
number of memory location points (# POINTS) in the sweep. When config-
ured to sweep past point 100, the sweep automatically wraps around to point 1.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals