13-12 Digital I/O Port, Safety Interlock, and Output Configuration 2400 Series SourceMeter
®
User’s Manual
Output configuration programming example
Table 13-3 lists the command sequence for output configuration. These commands set up
the SourceMeter as follows:
Interlock: enabled
Output-off mode: hi impedance
Auto-off mode: on
NOTE Connect pins 8 and 9 of the digital I/O port together to simulate a closed inter-
lock switch. Otherwise, the unit will not turn on its output when the measure-
ment is made.
Table 13-3
Output configuration programming example
Command Description
*RST
:SOUR:VOLT 10
:OUTP:INT:STAT ON
:OUTP:SMOD HIMP
:SOUR:CLE:AUTO ON
:READ?
Restore GPIB defaults.
Output 10V.
Enable interlock.*
Select high impedance output-off mode.
Enable auto-off mode.
Trigger and acquire readings.
*Connect pins 8 and 9 of digital I/O port to simulate closed interlock switch.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals