2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-61
[:ON] <function list>
OFF <function list>
[:SENSe[1]]:FUNCtion[:ON] <function list> Specify functions to be enabled
[:SENSe[1]]:FUNCtion:OFF <function list> Specify functions to be disabled
Parameters <function list> = “CURRent[:DC]” Amps measurement function
“VOLTage[:DC]” Volts measurement function
“RESistance” Ohms measurement function
NOTE Each function in the list must be enclosed in quotes (double or single) and sepa-
rated by a comma (i.e., :func:on “volt,” “curr”).
Query [:ON]? Query functions that are enabled
:OFF? Query functions that are disabled
Description When concurrent measurements are enabled, these commands are used
to enable or disable functions to be measured. The [:ON] command is
used to include (enable) one or more measurement functions in the list,
and the :OFF command is used to remove (disable) one or more func-
tions from the list.
Note that each function specified in the list must be enclosed in single or
double quotes, and functions must be separated by commas (,).
Examples:
:FUNCtion “VOLTage”, “CURRent” Enable volts and amps
functions
:FUNCtion:OFF ‘VOLTage’, ‘CURRent’ Disable volts and amps
functions
Note that there is a stand-alone command that can be used to enable or
disable all three measurement functions (see “ALL,” page 18-62).
If concurrent measurements (see “CONCurrent <b>,” page 18-60) are
disabled, the :ON command can only turn on one function at a time.
NOTE For the Model 2430 Pulse Mode, only one function can be enabled. Trying to
enable more than one function results in error -108.