18-116 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Query timestamp
TIME?
:SYSTem:TIME? Query timestamp
Query :TIME? Query timestamp
Description This query returns the current timestamp value.
Reset timestamp
RESet
:SYSTem:TIME:RESet Reset timestamp
Description This action command is used to reset the absolute timestamp to 0 sec-
onds. The timestamp also resets to 0 seconds every time the Source-
Meter is turned on.
Auto reset timestamp
RESet:AUTO <b>
:SYSTem:TIME:RESet:AUTO <b> Reset timestamp when exiting idle
Parameters <b> = 1 or ON Enable auto timestamp reset
0 or OFF Disable auto timestamp reset
Query :AUTO? Query enabled/disabled auto timestamp reset
state
Description :RES:AUTO enables or disables auto timestamp reset. When enabled,
the timestamp will be automatically reset when exiting the idle layer of
the trigger model. This command is intended for use with READ?/INIT
when taking more than one reading.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com