2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-81
2410
<n> = -1100 to 1100 Specify V-Source limit
20 Set limit to 20V
40 Set limit to 40V
100 Set limit to 100V
200 Set limit to 200V
300 Set limit to 300V
400 Set limit to 400V
500 Set limit to 500V
501 to 1100 Set limit to NONE
NONE Set limit to 1100V
DEFault Set limit to 1100V (NONE)
MINimum Set limit to 20V
MAXimum Set limit to 1100V (NONE)
2420
<n> = -63 to 63 Specify V-Source limit
6 Set limit to 6V
12 Set limit to 12V
18 Set limit to 18V
24 Set limit to 24V
30 Set limit to 30V
36 Set limit to 36V
48 Set limit to 48V
49 to 63 Set limit to NONE
NONE Set limit to 63V
DEFault Set limit to 63V (NONE)
MINimum Set limit to 6V
MAXimum Set limit to 63V (NONE)
2425 and 2430
<n> = -105 to 105 Specify V-Source limit
10 Set limit to 10V
20 Set limit to 20V
30 Set limit to 30V
40 Set limit to 40V
50 Set limit to 50V
60 Set limit to 60V
80 Set limit to 80V
81 to 105 Set limit to NONE
NONE Set limit to 105V
DEFault Set limit to 105V (NONE)
MINimum Set limit to 10V
MAXimum Set limit to 105V (NONE)