EasyManua.ls Logo

Keithley SourceMeter 2410

Keithley SourceMeter 2410
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2400 Series SourceMeter
®
User’s Manual Contact Check Function F-15
Command Description
:FORMat:ELEMents:CALCulate STATus Bit 18 (262144) of the status word returned by READ?,
FETCh?, MEASure?, MEASure:VOLTage?, MEA-
Sure:CURRent?, MEASure:RESistance?, TRACe:DATA?,
CALCulatel :DATA?, or :CALCulate2:DATA? signifies the
status of contact check (1 = open lead, 0 = OK). Also, bit 8 set
if LIMit4 is enabled.
:SYSTem:RSENse ON Enable 4-wire remote sense for contact check.
:SYSTem:RSENse? Query sense selection.
:SYSTem:CCHeck ON | OFF Enable/disable contact checking.
:SYSTem:CCHeck? Query state of contact check test.
:SYSTem:CCHEck:RESistance <NRf> Select threshold resistance (0 to 60):
<NRf> = 0 to 2 Selects 2 threshold resistance
= >2 to 15 Selects 15 threshold resistance
= >15 to 60 Selects 50 threshold resistance
:SYSTem:CCHeck:RESistance? Query threshold resistance.
:TRIGger:SEQuence2:SOURce <name> Enable or disable contact check event detection; <name> =
CCHeck (enable) or IMMediate (disable).
:TRIGger:SEQuence2:SOURce? Query state of event detection.
:TRIGger:SEQuence2:TOUT <NRf> Specify contact check timeout in seconds (0.00000 to
999.99990).
:TRIGger:SEQuence2:TOUT? Query contact check timeout setting.
*RST and SYSTem:PRESet parameter defaults are shown in boldface.
Table F-3 (continued)
Contact check remote commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com