134 AEJ-C/AES-C/PLJ-C-BA-e-1621
Printout example KERN YKB-01N
-------------------Pipette calibration---------------------
Meas. No.
10
Date
05.01.2016
Time
11:34:17
Temp.
20.0°C
Humidity
50 %
Pressure
1017hPa
-------------------------Tested volume 1000 µl------------
1
1003 µl
2
993 µl
3
1013 µl
4
1003 µl
5
993 µl
6
1003 µl
7
1013 µl
8
1053 µl
9
1003 µl
10
1023 µl
Average Volume [Va]
1010 µl
Arithmetic average
Systematic errorr [ES]
1.03 %
Systematic measurement deviation
Random error [Sr]
17.7 µl
Standard deviation
---------------------- Tested volume 5000 µl------------
1
4966 µl
2
4966 µl
3
4966 µl
4
4986 µl
5
4976 µl
6
4966µl
7
4966 µl
8
4976 µl
9
4976 µl
10
4976 µl
Average Volume [Va]
4972 µl
Arithmetic average
Systematic errorr [ES]
0.56 %
Systematic measurement deviation
Random error [Sr]
7.0 µl
Standard deviation
--------------------- Tested volume 10000 µl-------------
---
1
10033 µl
2
10033 µl
3
10033 µl
4
10033 µl
5
10043 µl
6
10043 µl
7
10043 µl
8
10043 µl
9
10043 µl
10
10043 µl
Average Volume [Va]
10039 µl
Arithmetic average
Systematic error [ES]
0.39 %
Systematic measurement deviation
Random error [Sr]
5.2 µl
Standard deviation
-------------------------------------------------------------------
Signature
…………………………………………………………