AEJ-C/AES-C/PLJ-C-BA-e-1621 133
2. Measuring sequence for test volume V2:
Follow the same sequence of operations for the second measuring sequence
but apply test volume <V2>wit
3. Measuring sequence for test volume V3:
Follow the same sequence of operations for the third measuring sequence but
apply test volume <V3>
After pressing the PRINT-key and the display of the systematic measurement
deviation [E
s
] / default deviation [S
r
] printing of log will take place via RS232
interface if an optional printer is connected (See sample printout below).
To start a new pipette test, press the F-key.
Return to weighing mode: