LXI:MDNS:SNAMe:DESired "name"
LXI:MDNS:SNAMe:DESired?
Sets the desired mDNS service name.
Parameter Typical return
Quoted string of up to 63 characters. "Keysight Technologies<model number> Data Acquisition System - <serial>", where
<serial> is the last 5 digits of the instrument's serial number.
Sets the desired mDNS service name:
LXI:MDNS:SNAM:DES "DAQ970A"
– This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
– This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
LXI:MDNS:SNAMe[:RESolved]?
The resolved mDNS service name will be the desired service name (LXI:MDNS:SNAMe:DESired command), possibly
with "(<N>)" appended. The N is an integer appended if necessary to make the name unique. The desired name may
be truncated, if necessary, to make room for the appended integer.
Parameter Typical return
(none) "Keysight Technologies DAQ970A Data Acquisition System - 00123"
Returns the resolved mDNS service name: LXI:MDNS:SNAMe:RESolved?
LXI:MDNS[:STATe] {OFF|ON}
LXI:MDNS[:STATe]?
Enables (On) or disables (Off) the multicast Domain Name System (mDNS), which provides the capabilities of a DNS
server for service discovery in a small network without a DNS server.
Parameter Typical return
{OFF|ON}
Default: ON
0 (OFF) or 1 (ON)
Enables mDNS:LXI:MDNS ON
– This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
– This parameter is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
Keysight DAQ970A/DAQ973A Programming Guide 161