LACO PROTOCOL DESCRIPTIONS TITANTEST™
117
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
12.4 Commands**
Command Name Data type R/W Meaning
dez hex
0 0000 NOP NO_DATA R No operation, replies without data
1 0001 Start NO_DATA W Switch to measure
2 0002 Stop NO_DATA W Switch to standby
3 0003 Vent NO_DATA W Switch to vent
4 0004 Calibration NO_DATA W Start, acknowledge calibration
5 0005 Clear error NO_DATA W Clear Error or Warning
6 0006 Zero UINT8 R/W 0 = Zero "Off"
1 = Zero "On" "Update"
without data = toggle Zero state
9 0009 Emission nominal status UINT8 R/W Emission nominal status
0 = OFF
1 = STANDBY
2 = ON
10 000A TMP nominal status UINT8 R/W TMP nominal status
0 = OFF
1 = ON
26 001A Interface protocol UINT8[2] R/W Interface protocol
Index 0: Remote Control
Index 1: RS232 / RS485
see table / enumerations
27 001B Used interface UINT8 R Used interface
0 = Remote Control
1 = RS232 / RS485
128 0080 Leak rate [sel. unit] FLOAT R Leak rate [sel. unit], limited to lower dis-
play limit
129 0081 Leak rate [mbar*l/s] FLOAT R Leak rate [mbar*l/s], unlimited
130 0082 Internal pressure 1 [sel. unit] FLOAT R Internal pressure 1 [sel. unit]
131 0083 Internal pressure 1 [mbar] FLOAT R Internal pressure 1 [mbar]
132 0084 Internal pressure 2 [sel. unit] FLOAT R Internal pressure 2 [sel. unit]
133 0085 Internal pressure 2 [mbar] FLOAT R Internal pressure 2 [mbar]
138 008A TMP actual rotation speed [Hz] UINT16 R TMP actual rotation speed [Hz]
139 008B TMP power [W] FLOAT R TMP power [W]
140 008C TMP operation hours [h] UINT32 R TMP operation hours [h]
141 008D Frequency converter operation hours [h] UINT32 R Frequency converter operation hours [h]
142 008E Leak detector operation hours [h] UINT32 R Leak detector operation hours [h]
143 008F TMP temperature bottom [deg. C] FLOAT R TMP temperature bottom [deg. C]
144 0090 TMP temperature electronic [deg. C] FLOAT R TMP temperature electronic [deg. C]
145 0091 TMP temperature bearing [deg. C] FLOAT R TMP temperature bearing [deg. C]
146 0092 TMP temperature motor [deg. C] FLOAT R TMP temperature motor [deg. C]
** Feature not available in TITANTEST™ Maintenance model