LACO PROTOCOL DESCRIPTIONS TITANTEST
118
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
147 0093 Time since power on [min] UINT32 R Time since power on [min]
150 0096 TMP voltage [V] FLOAT R TMP voltage as reported by TMP control-
ler [V]
151 0097 TMP current [A] FLOAT R TMP current as reported by TMP control-
ler [A]
158 009E TMP runup time [s] UINT16 R TMP runup time [s]
159 009F Time in measure [s] UINT16 R Time in measure [s]
160 00A0 Time in evacuation [s] UINT16 R Time in evacuation [s]
161 00A1 Time in standby [s] UINT16 R Time in standby [s]
165 00A5 Electronic temperature [deg. C] FLOAT R Electronic temperature [deg. C]
166 00A6 Preamplifier temperature [deg. C] FLOAT R Preamplifier temperature [deg. C]
167 00A7 Anode voltage [V] FLOAT R Anode voltage [V]
168 00A8 Cathode voltage [V] FLOAT R Cathode voltage [V]
169 00A9 Suppressor voltage [V] FLOAT R Suppressor voltage [V]
170 00AA Anode-cathode voltage [V] FLOAT R Anode-cathode voltage [V]
188 00BC Maximum leak rate [sel. unit] FLOAT R Maximum leak rate since last inquiry via
interface [selected unit]
189 00BD Maximum leak rate [mbar*l/s] FLOAT R Maximum leak rate since last inquiry via
interface [mbar*l/s]
202 00CA Pre amplifier voltage [V] FLOAT R Pre amplifier voltage [V]
212 00D4 24 V power out RC [V] FLOAT R 24 V power out RC [V]
213 00D5 24 V power out IO [V] FLOAT R 24 V power out IO [V]
221 00DD Analog outputs [V] FLOAT[2] R Analog outputs [V]
Index 0: Channel 1
Index 1: Channel 2
222 00DE Analog output configuration UINT8[2] R/W Function of analog output
Index 0: Channel 1
Index 1: Channel 2
see table / enumerations
223 00DF Analog output leak rate scale (log. only UINT8 R/W Leak rate scaling of analog output in log-
arithmic mode
see table / enumerations
224 00E0 Analog output upper exponent SINT8 R/W Upper limit for the analog out at the I/O
modul. Value is exponent of the mbar*l/
s value. Example: -5 = 1E-5 mbar*l/s
260 0104 Calibration status UINT8 R Status of calibration
see table / enumerations
261 0105 PLC input state UINT16 R Get PLC input state
see table / enumerations
262 0106 PLC output state UINT16 R Get PLC output state
see table / enumerations
Command Name Data type R/W Meaning
dez hex