LACO PROTOCOL DESCRIPTIONS TITANTEST™
124
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
602 025A Audio alarm delay [s] UINT16 R/W Audio alarm delay
0, 1, 2, 3, ...
10, 12, 14, ....
30, 35, 40, ...
100, 110, 120,
300, 330, 360, ... 600
60000 = infinite
604 025C Audio beep UINT8 R/W Audio beep
0 = disabled
1 = enabled
626 0272 Bypass evacuation UINT8 R/W Bypass option evacuation
0 = Fore pump only
1 = Both pumps
2 = External pump only
627 0273 Bypass measure UINT8 R/W Bypass option measure
0 = Fore pump only
1 = Both pumps
2 = External pump only
630 0276 AutoTest delay timer [s] UINT16 R/W Auto Test Delay timer [s]
631 0277 AutoTest test timer [s] UINT16 R/W Auto Test test timer [s]
0 means infinite
632 0278 AutoTest background timer [s] UINT16 R/W AutoTest background timer [s]
0 means infinite
640 0280 AutoTest background [mbar*l/s] FLOAT R/W AutoTest background limit [mbar*l/s]
641 0281 AutoTest external trigger UINT8 R/W AutoTest external trigger
0 = disabled
1 = enabled
642 0282 AutoTest reject mode UINT8 R/W AutoTest reject mode
0 = Q max
1 = Q last
645 0285 Fill complete NO_DATA W Autotest Fill complete
650 028A Postamplifier range UINT8 R Postamplifier range
0 = 0.4
1 = 1.6
2 = 6.4
3 = 25.6
4 = 0.4 (FIXED)
5 = 1.6 (FIXED)
6 = 6.4 (FIXED)
7 = 25.6 (FIXED)
652 028C Postamplifier voltage [V] FLOAT R Postamplifier voltage [V]
700 02BC AutoTest status UINT8 R AutoTest status
see table / enumerations
Command Name Data type R/W Meaning
dez hex