LACO PROTOCOL DESCRIPTIONS TITANTEST
123
© 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807
504 01F8 500GOhm value FLOAT R/W 500GOhm value
506 01FA Mass UINT8 R/W Mass
2 = Mass 2 (H2)
3 = Mass 3
4 = Mass 4 (Helium)
515 0203 Set desired leak rate FLOAT W Set desired leak rate (KnL)
517 0205 Offset internal [A] FLOAT[3] R Offset internal
Index 0: mass 2
Index 1: mass 3
Index 2: mass 4
525 020D Calibration factors FLOAT[18] R/W Calibration factors
see table / enumerations
529 0211 Warnings UINT8 R/W Warnings
0 = disabled
1 = enabled
530 0212 Cathode selection UINT8 R/W Cathode selection
1 = CAT 1
2 = CAT 2
540 021C Pressure EVAC GROSS [mbar] FLOAT R/W Pressure limit EVAC --> GROSS [mbar]
541 021D Pressure GROSS FINE [mbar] FLOAT R/W Pressure limit GROSS --> FINE [mbar]
543 021F Pressure FINE ULTRA [mbar] FLOAT R/W Pressure limit FINE --> ULTRA [mbar]
548 0224 Background subtraction UINT8 R/W Background subtraction
0 = disabled
1 = enabled
550 0226 Vacuum ranges UINT8[3] R/W Vacuum ranges
Index 0: GROSS
Index 1: FINE
Index 2: ULTRA
553 0229 Vent mode UINT8 R/W Vent mode
0 = Vent with Stop
1 = Vent manual
2 = No Vent
555 022B Evacuation time [s] UINT16 R/W Evacuation time [s]
600 0258 Audio alarm type UINT8 R/W Audio alarm type
0 = Leak rate prop
1 = Trigger alarm
2 = Setpoint
3 = Pinpoint
Command Name Data type R/W Meaning
dez hex